These pyrocoated HGA graphite tubes with integrated platform, pack of 5, fit the HGA graphite furnace of the PinAAcle® 900H atomic absorption spectrometers.
Graphite tubes with integrated platforms are machined from one piece of graphite and totally pyrocoated, for exceptional tube-to-tube reproducibility; the spherically shaped platform takes up to 50 µL of sample, provides STPF conditions for volatile and non-volatile refractory elements through the higher heating rates of tube and platform, saves inserting a separate platform, and gives longer life with aggressive matrices such as concentrated acids.
PerkinElmer makes all its graphite tubes from a specific grade of high-density base graphite developed for its exclusive use, because the composition and microstructure of the base material set the tube’s resistivity and heating characteristics; all graphite parts, contacts, tubes and platforms, are held to strict quality control so that analytical conditions stay stable from one atomization cycle to the next and from tube to tube.
Each manufacturing lot of PerkinElmer graphite tubes is tested to specifications for lifetime, sensitivity and precision after mechanical, electrical, thermal and analytical testing, and the pyrolytic coating is applied to a specified thickness because it sets the tube’s resistance and temperature distribution.
Every tube carries a dot mark that shows the orientation of the platform for correct installation, and new tubes are conditioned before use with the Condition Tube program in Syngistix™ for AA.
Product Compatibility | |
Brand Name | PerkinElmer |
Product Group | |
Style | Integrated Platform |
Technology Type |