These UltraClean THGA graphite tubes with integrated platform, pack of 20, fit the transversely heated graphite furnace of the PinAAcle® 900T and 900Z atomic absorption spectrometers, for analyses where the residual contaminant levels of standard tubes are not low enough.
UltraClean tubes are cleaned by extended high-temperature gas-phase procedures and are tested and specified for blank levels of Cr, Ni, V, Mo and Ti below 0.003 integrated absorbance units and Na, Fe, Al, Si and Ca below 0.0005, for the lowest detection limits in applications such as ultra-trace Na, Ca, Fe, Cr and Ni in semiconductor reagents and ultra-trace environmental, clinical and food analyses.
PerkinElmer’s transversely heated graphite tube carries an integrated platform; the exceptionally uniform temperature distribution of transverse heating reduces or eliminates condensation of sample matrix components and memory effects and improves atomization efficiency for refractory elements.
PerkinElmer makes all its graphite tubes from a specific grade of high-density base graphite developed for its exclusive use, because the composition and microstructure of the base material set the tube’s resistivity and heating characteristics; all graphite parts, contacts, tubes and platforms, are held to strict quality control so that analytical conditions stay stable from one atomization cycle to the next and from tube to tube.
Each manufacturing lot of PerkinElmer graphite tubes is tested to specifications for lifetime, sensitivity and precision after mechanical, electrical, thermal and analytical testing, and the pyrolytic coating is applied to a specified thickness because it sets the tube’s resistance and temperature distribution.
Every tube carries a dot mark that shows the orientation of the platform for correct installation, and new tubes are conditioned before use with the Condition Tube program in Syngistix™ for AA.
Product Compatibility | |
Brand Name | PerkinElmer |
Product Group | |
Style | Integrated Platform |
Technology Type |