These Advanced Platform THGA graphite tubes with end caps, pack of 5, fit the transversely heated graphite furnace of the PinAAcle® 900T and 900Z atomic absorption spectrometers.
The patented Advanced Platform suspends the platform in the tube on three small pins in a triangular layout instead of attaching it along one side; this minimises wall contact and thermal conduction from the tube walls, so the platform heats evenly and rapidly, almost entirely by thermal radiation, which gives improved sensitivity for some applications and improved batch-to-batch reliability.
The end-cap tubes are made for improved characteristic mass with volatile and medium-refractory elements such as Ag, As, Cd, Co, Cr, Cu, Pb, Se and Tl, giving improved detection limits for those elements and faster drying of complex matrices through the enlarged dosing hole; they are not recommended for refractory elements such as V or Ba.
PerkinElmer makes all its graphite tubes from a specific grade of high-density base graphite developed for its exclusive use, because the composition and microstructure of the base material set the tube’s resistivity and heating characteristics; all graphite parts, contacts, tubes and platforms, are held to strict quality control so that analytical conditions stay stable from one atomization cycle to the next and from tube to tube.
Each manufacturing lot of PerkinElmer graphite tubes is tested to specifications for lifetime, sensitivity and precision after mechanical, electrical, thermal and analytical testing, and the pyrolytic coating is applied to a specified thickness because it sets the tube’s resistance and temperature distribution.
Every tube carries a dot mark that shows the orientation of the platform for correct installation, and new tubes are conditioned before use with the Condition Tube program in Syngistix™ for AA.
- Other pack sizes: pack of 10 N9304645, pack of 20 N9307833
- Advanced Platform tubes without end caps: pack of 5 N9307830, pack of 10 N9304644, pack of 20 N9307831; standard tubes with end caps pack of 5 B3000653, pack of 10 N9304643, pack of 20 B3000655
- THGA contact cylinders B0504035 and B0504036; quartz window B0198005; inspection mirror B0851900
Product Compatibility | |
Brand Name | PerkinElmer |
Product Group | |
Style | Advanced Platform |
Technology Type |