These pyrocoated HGA graphite walled tubes, pack of 10, fit the HGA graphite furnace of the PinAAcle® 900H atomic absorption spectrometers. The pyrocoating gives improved sensitivity and tube-to-tube reproducibility, and the walled tube is used in place of platform tubes when analysing refractory elements such as vanadium, molybdenum and titanium; it is not compatible with L’vov platforms.
PerkinElmer makes all its graphite tubes from a specific grade of high-density base graphite developed for its exclusive use, because the composition and microstructure of the base material set the tube’s resistivity and heating characteristics; all graphite parts, contacts, tubes and platforms, are held to strict quality control so that analytical conditions stay stable from one atomization cycle to the next and from tube to tube.
Each manufacturing lot of PerkinElmer graphite tubes is tested to specifications for lifetime, sensitivity and precision after mechanical, electrical, thermal and analytical testing, and the pyrolytic coating is applied to a specified thickness because it sets the tube’s resistance and temperature distribution.
Every tube carries a dot mark that shows the orientation of the platform for correct installation, and new tubes are conditioned before use with the Condition Tube program in Syngistix™ for AA.
Product Compatibility | |
Brand Name | PerkinElmer |
Product Group | |
Style | Walled |
Technology Type |